DELTA LTL2000 Retrometer 17
é to Change Press é to disable <ENTER> to accept.
Clear Top Log
(Activate with the <é> key after a Rl measurement)
LCD Display Explanation.
Clear Top Log?
Enter to Confirm
Removes the newest measurement from the Rl Data Log.
Confirm the removal with the <ENTER> key.
Clear Top Log?
No Data in Log!
This message is shown when the Rl Data Log is empty.
Rl:..mcd/m
2
/lx
(ID-Text) or time
The previous Rl result will be shown and if this measurement has an
ID then this ID will also be restored together with its serial number.
The time will not be updated!
Clear Top T-Log
(Activate with the <é> key after a Test, Calib or Zero measurement)
LCD Display Explanation.
Clear Top T-Log?
Enter to Confirm
Removes the newest measurement from the Test Log.
Confirms the removal with the <ENTER> key.
Clear Top T-Log?
No Data in Log!
This message is shown when the Test Log is empty.
Rl:....mcd/m
2
/lx
time
The previous result will be shown. The time will not be updated!
Measurement ID
(Activate with the <ê> key)
LCD Display Explanation.
Seq ID:ABcd12
Enter to Edit
The Measurement ID is shown. Press the <ENTER> key in order to
edit the 6 position ID string.
Seq ID:ABcd12
Edit é______
An é indicates the active edit position, use <é><ê> to change the
value above the pointer. Press <ENTER> to switch to the next
position.
Seq ID:AB__#1
Edit ______é
Press <ENTER> on the last position to activate the new
Measurement ID, this also resets the serial number to 0 for the new
ID.
Vista de pagina 22