Operator's Manual
Digital Setup
Digital Logic Setup
Each lead bank requires a Logic Setup to determine the assignment of a 1 or 0, depending
on the measured voltage relative to the Threshold.
Threshold Level
The threshold level determines how the input signal is interpreted. Input voltages less than
the threshold are converted to 0, while input voltages greater than the threshold are
converted to 1.
To set Threshold:
1. Go to Vertical > Digital 1 Setup to Display the Digital Setup dialog. Open the Logic
Setup tab.
2. For each available lead bank, Touch Logic Family and either chose one of the
standard families or User Defined.
TTL circuits use a threshold voltage level of 1.58 V. ECL circuits use a threshold
voltage level of -1.39 V.
3. If you chose User Defined, enter a custom Threshold and Hysteresis.
Custom threshold levels can be set between –10.0 V and +10.0 V.
13